Vista de página 118
GS-800 Imaging Densitometer
7-5
Choosing Your Own Settings
If you know the filter and light source settings you want, or want to
experiment with different settings, you can choose them yourself.
Fig. 7-4. GS-800 Custom Application controls.
Next to Filter, click on either the Red, Green, Blue, or White checkbox, or a
combination of two of the first three (Red-Green, Green-Blue, Red-Blue).
Fig. 7-5. Examples of filter colors and applications.
Next to Light, select the appropriate light source.
Reflective mode is for scanning opaque mediums such as dried gels on
filter paper, arrays, or photographs. The scanning dimensions in this
mode are 21 cm x 29 cm.
Transmissive mode is for scanning transparent mediums such as films,
gels, or slides. The scanning dimensions in this mode are 20 cm x 23 cm.
Filter Color Wavelength Application Examples
Red 595–750 nm Coomassie G-250, BCIP/NBT, Fast Green
FCF, Methylene Blue
Green 520–570 nm Coomassie R-250, Basic Fuchsin
Blue 400–530 nm Crocein Scarlet
Gray Scale 400–750 nm Silver Stains, Copper Stains, Film